Implementation of the ABL-90 blood gas analyzer in a ground-based mobile emergency care unit

Abstract

Point-of Care analysis is increasingly being applied in the prehospital scene. Arterial blood gas analysis is one of many new initiatives adding to the diagnostic tools of the prehospital physician. In this paper we present a study on the feasibility of the Radiometer ABL-90 in a ground-based Mobile Emergency Care Unit and report on some clinical situations in which the apparatus has proven beneficial.

Point-of-Care analysis and treatment is increasingly being applied in the prehospital setting. Obtaining an electrocardiogram and initiating thrombolysis is by now an established part of the prehospital treatment in coronary occlusion [1]. Increasingly, ultrasound is applied prehospitally [2, 3] and prehospital computerized tomography of the head and ensuing thrombolysis in case of stroke is probably approaching as a treatment modality [4]. Prehospital analysis of blood samples for selected parameters is already applied in some circumstances, most notably measurement of blood glucose, but also troponin and lactate are measured prehospitally [5, 6].

Fig. 1
figure 1

The ABL-90 is secured in its cradle situated to the left of the observers seat behind the physician´s seat

However, full blood gas analysis in prehospital services requires a more extensive set-up than previously described. Within the hospital, the ABL-90 (Radiometer, Denmark) blood gas analyzer has proven fast and reliable. The ABL-90 is portable and data can be sent to the patient data management systems. An internal battery provides back-up when disconnected from the inverter. The apparatus thus would appear suitable for prehospital use.

We report a study on the feasibility of the ABL-90 implemented in the anaesthesiologist staffed Mobile Emergency Care Unit (MECU) in Odense, Denmark.

Initially, the vehicle was equipped with an inverter capable of supplying the ABL-90 with 90 Watts and 230 Volts alternating current. Following that, the susceptibility of the ABL-90 towards vibrations and gravitational forces was assessed. A dummy with dimensions and weight identical to the ABL-90 was constructed and mounted in the vehicle. An accelerometer was placed on the dummy, measuring g-forces during several emergency MECU runs. Furthermore, the shock response spectrum of the unit was calculated. The extent of g-forces exerted and the shock response spectrum measured confirmed the possible functionality of the ABL-90 following the physical impact endured during the emergency runs, and further testing was carried out.

Although g-forces exerted were rather low (1.5–1.95 g in all directions), initial findings demonstrated that the ABL-90 was unreliable. An inlet of the apparatus proved to be responsible for failure, disconnecting the inlet from the calibration system. A re-designed cradle with added cushioning and a newly constructed sample inlet, designed to minimize the risk of a discontinued connection between the sample inlet and solution pack corrected this issue enabling the ABL-90 to function flawlessly. In order to ensure correct results, validation of the blood sample analyses was performed by the Department of Clinical Biochemistry at Odense University Hospital. A maintenance program was also established ensuring that a supplementary ABL-90 is always accessible should maintenance or replacement of solution packs require the unit to be disassembled.

For a period of 18 months the ABL-90 has been functioning in the MECU in Odense, Denmark Fig 1.

The ability to perform rapid, accurate arterial blood gas analysis at the scene has significantly improved the diagnostic possibilities of the MECU, as the diagnosis of several conditions in which a fast diagnosis and subsequent therapy is vital, thus has been made possible.

Among many clinical scenarios where arterial blood gas analyses are performed at the scene, some of the special conditions where the results have proven relevant have been:

In patients with presumed septic shock or septicaemia, analysis of lactate and pH in many cases has enabled the MECU to establish the diagnosis prehospitally, to obtain blood cultures and to initiate antibiotic therapy at the scene.

In several cases, patients suspected of exposure to Carbon monoxide poisoning have been released at the scene thus obviating admission to hospital once blood gas analysis has rejected the presence of Carboxyhemoglobin.

In other related cases, patients exposed to indoor fires with potential for cyanide formation have been diagnosed as the shift to anaerobic metabolism caused by cyanide has been detected by lactic acidosis.

In patients with exacerbation of chronic obstructive pulmonary disease, rapid blood gas analysis has helped the physician at the scene to better direct patient therapy.

Also, the blood gas analyser is helpful in adjusting the level of respiratory support in intubated patients with elevated intracranial pressure, in which the CO2-level should be targeted within rather narrower limits than interpretation of end-tidal CO2 measurement allows.

As the apparatus has the potential of transmitting prehospital results to intra-hospital patient data management systems it facilitates improved diagnostic and treatment possibilities, both prehospitally and in preparing emergency departments for patients before their arrival at the departments. Transmission of data to the receiving emergency department enables these departments to increase their levels of preparedness and thereby initiate the correct treatment faster.

The direct costs associated with the technology are acquisition of the apparatus. The retail price is EUR 15.000 per unit, while indirect costs (including reagents, syringes etc.) amount to EUR 11.500 per year. Implementation is relatively straightforward. Only minute changes have been made in the MECU to make the ABL-90 suitable for the unit.

Conclusion

We have successfully introduced the possibility of acquiring full blood gas analyses prehospitally in a mobile emergency care unit.

References

  1. Peterson MC, Syndergaard T, Bowler J, Doxey R. A systematic review of factors predicting door to balloon time in ST-segment elevation myocardial infarction treated with percutaneous intervention. Int J Cardiol. 2012;157(1):8–23. doi:10.1016/j.ijcard.2011.06.042. Epub 2011 Jul 14.

    Article  PubMed  Google Scholar 

  2. Foss KT, Subhi Y, Aagaard R, Bessmann EL, Bøtker MT, Graumann O, et al. Developing an emergency ultrasound app – a collaborative project between clinicians from different universities. Scand J Trauma Resusc Emerg Med. 2015;23:47. doi:10.1186/s13049-015-0130-2. Published online 2015 Jun 20.

    Article  PubMed Central  PubMed  Google Scholar 

  3. Nelson BP, Sanghvi A. Out of hospital point of care ultrasound: current use models and future directions. Eur J Trauma Emerg Surg. 2015. doi:10.1007/s00068-015-0494-z. Epub ahead of print.

    PubMed  Google Scholar 

  4. Hov MR, Nome T, Zakariassen E, Russell D, Røislien J, Lossius HM, et al. Assessment of acute stroke cerebral CT examinations by anaesthesiologists. Acta Anaesthesiol Scand. 2015. doi:10.1111/aas.12542 [Epub ahead of print].

    PubMed  Google Scholar 

  5. Stengaard C, Sørensen JT, Ladefoged SA, Christensen EF, Lassen JF, Bøtker HE, et al. Quantitative point-of-care troponin T measurement for diagnosis and prognosis in patients with a suspected acute myocardial infarction. Am J Cardiol. 2013;112(9):1361–6. doi:10.1016/j.amjcard.2013.06.026. Epub 2013 Aug 14.

    Article  CAS  PubMed  Google Scholar 

  6. Guyette FX, Meier EN, Newgard C, McKnight B, Daya M, Bulger EM, et al. A comparison of prehospital lactate and systolic blood pressure for predicting the need for resuscitative care in trauma transported by ground. J Trauma Acute Care Surg. 2015;78(3):600–6. doi:10.1097/TA.0000000000000549.

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Søren Mikkelsen.

Additional information

Competing interests

The authors declare that they have no conflicts of interest. No external funding was acquired.

Authors´ contributions

SM, JWH, CM and SJ conceived the original idea. JWH and CM performed the vibrational and gravitational measurements. MN facilitated the maintenance program and was responsible for the validation process. SM, MN and SJ drafted the manuscript. All authors read and approved the final manuscript.

Rights and permissions

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Mikkelsen, S., Wolsing-Hansen, J., Nybo, M. et al. Implementation of the ABL-90 blood gas analyzer in a ground-based mobile emergency care unit. Scand J Trauma Resusc Emerg Med 23, 54 (2015). https://doi.org/10.1186/s13049-015-0134-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13049-015-0134-y

Keywords